Balanced Adjustment (Zero Adjustment)(For more information, please refer to the instruction manual “Zero-point adjustment of measurement position”.)

When strain is measured with the strain measurement unit, balance adjustment function can be used, to set the measurement value before measurement to zero.
Please set measurement unit in “Settings -> Setup…” and execute checking disconnection before completing establishment of communication for all units.
In case of executing high-speed measurement with the sampling period 50 ms or less, establishment of communication for the high-speed data receiver is also needed.

Procedure

Connect the sensors for strain gauge or strain logger type to measurement unit, and conduct setting for standard state.
Click on “[Zero Point Adjustment…]” button in Setting.

Insert check marks according to each measurement unit and click on “Execute” button to execute the measurement.
When succeeding, “Status” is turned on green color and “Adjusted” is displayed.
When failing by error of communication, “Status” is turned on red color and “Error” is displayed.
If balanced adjustment is out of its range or disconnection occurs, “Status” is turned on red color and “Out Of Range” is displayed.